4200-42100 pages
SERVOMEX 4200M210
nnnnGAS PURITY ANALYSER
nnnnis specifically designed to monitor flammable gas
nnnnsamples including "HyCO" or "Syngas" mixtures
nnnnMeasures up to four gas streams simultaneously
nnnnSuitable for use with flammable samples +
nnnnEasy and quickly configured using intuitive LCD
nnnnmenu
nnnnLow maintenance
nnnnExtremely stable and reliable sensors
nnnnDigital comms: RS232/RS485, Modbus™ protocol
nnnnv c
nnnnSPECIFICATION
nnnnGases Measured:
nnnn02 (control)
nnnnC02 (%)'
nnnnco (%y
nnnnCH4 (%y
nnnnPERFORMANCE
nnnnTechnology:
nnnnMax Range:
nnnnMin Range:
nnnnIntrinsic error (accuracy):
nnnnZero drift/week2 :
nnnnSpan drift2 :
nnnnParamagnetic
nnnn0-100% 02
nnnn0-5% 02
nnnn<0.15% 02
nnnn<0.05% 02
nnnn<0.1% 02/week
nnnnSBSW Infrared «
nnnn0.25, 0.5, 1, 2.5, 5,
nnnn10,25, 50,100% CO
nnnn2
nnnn1 % of FS
nnnn<2% of FS
nnnn<1% of reading/day
nnnnSBSW Infrared °
nnnn1, 2.5, 10% CO
nnnn1 % of FS
nnnn<2%of FS
nnnn<1 % of reading/day
nnnnSBSW Infrared °
nnnn5, 25, 50, 100% CH«
nnnn1 % of FS
nnnn<2% of FS
nnnn<1% of reading/day
nnnnGases
nnnnMeasured:
nnnnC02 (trace)'
nnnnCO (trace)'
nnnnCH4 (trace)'
nnnnPERFORMANCE
nnnnTechnology
nnnnMax Range:
nnnnMin Range:
nnnnIntrinsic error (accuracy):
nnnnZero drift/week2 :
nnnnSpan drift/week2 :
nnnnGfx Infrared +
nnnn0-100ppm(v) C02«
nnnn0-5ppm(v) C02
nnnn<0.1ppm(v) C02**
nnnn<0.2ppm(v) C02
nnnn<2% of reading
nnnnor 0.2ppm(v) C02*
nnnnGfx Infrared +
nnnn0-500ppm(v) CO •
nnnn0-10ppm(v) CO
nnnn<0.5ppm(v) CO**
nnnn<1 ppm(v) CO
nnnn<2% of reading
nnnnor 1 ppm(v) CO*
nnnnGfx Infrared +
nnnn0-500ppm(v) CH4«
nnnn0-10ppm(v) CH4
nnnn<0.5ppm(v) CH4**
nnnn<1 ppm(v) CH4
nnnn<2% of reading
nnnnor 1 ppm(v) CH4*
nnnn• Higher ranges available, consult Servomex
nnnn+ gas filter correlation
nnnn© single beam single wavelength
nnnnor 1 % of reading whichever is greater
nnnnwhichever is greater
nnnnModbus™ is a trademark of AEG-MODICON
nnnnThe 4200 is used with general flammable samples (and
nnnns fitted with viton tubing). The 4210 is designed for use
nnnnwith samples containing more than 1% hydrogen or
nnnn0.5% acetylene. The 4210 has 316SS metal pipework
nnnnthroughout. Both versions should be installed in a
nnnnfreely ventilated non-hazardous area
nnnn1. Standard factory calibration is in a background of nitrogen. Consult Servomex for other background gases
nnnn2. Warm up time typically 24 hours from switch on. May be longer for some high sensitivity measurements
nnnnSERVOMEX"
"