Thermal Conductivity & Galvanic Gas Analyser (Panel Mount) - KG15500 pages
500-0009-1
KG1550 dual-sensor gas
analyser
Features
♦ Measures oxygen plus
CO2/H2/Ar/He/Freon etc
♦ Programmable alarms and
analogue output for each gas
♦ Microprocessor controlled
♦ Compact DIN panel-mounting
enclosure
The Hitech KG1550 analyser
tually no maintenance costs. By
The user can configure the alarm
combines accurate measurement
applying the power of a micro-
function (high, low or off) as well
of oxygen in mixtures that contain
processor to temperature com-
as the hysteresis value.
another constituent suitable for
pensation and linearisation, an
The instrument is packaged in
measurement by thermal conduc-
accuracy and measuring range
a standard DIN enclosure (96 x
tivity. Other gases measured in-
not
144mm) with an optional locking
clude carbon dioxide, helium and
katharometers is provided.
argon, giving the analyser a wide
suitable gases can be measured
For sampling, a variety of sys-
range of applications in both food
over 0 to 100%, with the maxi-
tems is available for any neces-
and process industries.
mum
sary pumps, filters and take-off
usually
associated
sensitivity
being
with
All
gas-
dependent, and ranges from 0 to
Oxygen is measured by an
tubes can also be supplied.
10% for carbon dioxide to 0 to
electrochemical cell that is virtu-
cover with a rating of IP54.
5000ppm for hydrogen.
ally maintenance-free.
The oxy-
Typical Applications
gen concentration range can be
Measured values are presented
from 0 to 100%, to 0 to 200ppm
on a large, clear LCD screen
dependent upon the other con-
which also displays messages
stituents of the gas mixture.
and prompts for the menu-driven
configuration and calibration rou-
The second gas is measured by
tines.
a non-depleting, katharometer
thermal conductivity cell which
Two concentration alarms are
combines high stability and low
provided for each gas and give
mass to form a sensor with fast
visual (LED) indication and volt-
response, high sensitivity and vir-
free changeover contact outputs.
♦ Food packaging
♦ Produce and fruit storage
♦ Hydrogenation processes
♦ Fermentation processes
♦ General process plant