SLFA-60 X-ray Fluorescence Sulfur-in-Oil Analyzer0 pages
Technical Specifications
Measurement Specifications | |
Measurment principle | Energy Dispersive X-ray Fluorescence Analysis Method (EDXRF) |
Measurement sample | Sulfur in Petroleum products such as heavy oil, naphtha, and crude oil, light oil |
ivicdouiciMCML idiiytr | D - Q QQQQ % |
Repeatability | 15 ppm or less (with a 1 % sulfur sample) |
C/H error correction | ± 50 ppm or less per C/H (with a 1 % sulfur sample) |
Lower detection limit | 20 ppm or less (three times the standard deviation with a 0 % sample) |
Number of calibration curves | 5 lines x 3 sets (Total 15 lines) |
Calibration curve order | Linear or quadratic (both automatic selection and manual settings available) |
Calibration | Calibration made using calibration points: (5 - 20 points) |
Sample requirement | 4 -10 ml |
Measurement time | 10 - 600 sec |
Spectrum measurement | Energy Axis: 0-10 keV, spectrum analysis to verify instrument preformances. |
External output | USB connect with PC, USB memory |
Ambient Conditions | 1 | |
Temperature | + 5 "C ~ + 40 "C (+ 41 °F ~ + 104°F) | |
Humidity | 80% max relative humidity in temperature range + 5 'C ~ + 30 'C (+ 41 °F ~ + 86°F). Linear decrease to 50% relative humidity in temperature range + 31 'C ~ + 40 'C (+ 88°F ~ + 104°F). |
Power and Housing | ||
Power supply | AC 100 - 240 V ± 10 %, 50/60 Hz | |
Power consumption | 80 VA | |
Dimentions [WxDxH] | 230x400x 140 mm (9x 15.75x5.5 in.) | |
Weight | 9 kg (20 lbs.) |
Comformity Standards | |
ASTM D4294(USA)-ISO 8754-JIS K2541/B7995 |
Dimentional Outlines (Unit
mm
£= °L
230
400
Please read the operation manual before using this product to assure safe and proper handling of the product.
•The contents of this catalog are subject to change without prior notice, and without any subsequent liability to this company.
•The color of the actual products may differ from the color pictured in this catalog due to printing limitations.
• It is strictly forbidden to copy the content of this catalog in part or in full.
•All brand names, product names and service names in this catalog are trademarks or registered trademarks of their respective companies.
•Windows is a registered trademark of Microsoft Corporation in the United States and other countries.
http://www.horiba.com e-mail: info@horiba.co.jp
• HORIBA, Ltd.
Head Office
2 Miyanohigashi, Kisshoin
Minami-ku, Kyoto, Japan
Phone: 81 (75) 313-8123
Fax: 81 (75) 321-5725
• HORIBA Thailand Ltd
• HORIBA (China) Trading Co., Ltd.-
Shanghai Office Beijing Office
Unit D, 1F, Building A, Synnex Room 1801, SKTower.
International Park, 1068
West Tianshan Road,
Shanghai, 200335 China
Phone: 86 (21) 6289-6060
Fax: 86 (21) 6289-5553
Tower 1 No.6 Jia
Jianguomenwai Ave.,
Chaoyang District, Beijing
100022 China
Phone: 86 (10) 8567-9966
Fax: 86 (10) 8567-9066
• HORIBA Instruments (Singapore) Pte Ltd.
393, 395, 397, 399, 401, 403 Head Office
Latya Road, Somdetchaopraya 10, Ubi Crescent #05-12
Sub-district Klongsan
District Bangkok 10600
Phone: 66 (2) 861 5995
Fax: 66 (2) 861 5200
• HORIBA India Private
Limited
Delhi Office
246, Okhla Industrial Estate,
Phase 3 New Delhi - 110020.
ndia
Phone: 91 (11) 4646-5000
Fax: 91 (11) 4646-5020
Lobby B Ubi Techpark
Singapore 408564
Phone: 65 (6) 745-8300
Fax: 65 (6) 745-8155
Hanoi Office
Unit 10, 4 Floor, CMC tower.
Duy Tan Street, Dich Vong
Hau Ward, Cau Giay district.
Hanoi, Vietnam
Phone: 84 (4) 3795-8552
Fax: 84 (4) 3795-8553
Guangzhou Branch
Room 1611/1612, Goldion
Digital Network Centar,
138Tiyu Road East,
Guangzhou, 510620 China
Phone: 86 (20) 3878-1883
Fax: 86 (20) 3878-1810
• HORIBA Korea Ltd.
10, Dogok-Ro, 6-Gil,
Gangnam-Gu, Seoul.
135-860, Korea
Phone: 82 (2) 753-7911
Fax: 82 (2) 756-4972
Jakarta Office
Menara Bidakara 2 Unit 11-04.
Jl. Jend. Gatot Subroto
Kav. 71-73, Jakarta
Selatan, 12870, Indonesia
Phone: 62 (21) 2906-9419
Fax: 62 (21) 2906-9421
• HORIBA Instruments Incorporated
Edison Office
3880 Park Ave.
Edison, NJ 08820, U.S.A
Phone: 1(732)494-8660
Fax: 1(732)549-5125
• HORIBA JOBIN WON SAS-
16-18, rue du Canal,
91165 Longjumeau Cedex.
Phone: 33 (0) 1 69 74 72 00
Fax: 33 (0) 1 69 09 07 21
HORIBA JOBIN WON GmbH
Haupt-Str. 1,82008
Unterhaching, Germany
Phone: 49 (89) 462 3170
Fax: 49 (89) 462 31799
HORIBA JOBIN WON Sri
Via Cesare Pavese 19/21,
200900 Opera, Mirano, Italy
Phone: 39 (2) 5760 3050
Fax: 39 (2) 5760 0876
Spain Branch
PAE Neisa Norte Ed. II
Avenida Vadelaparra,
27/28108 Alcobendas/
Madrid Spain
Phone: 34(91) 661 7987
Fax: 34(91)661 6658
Irvine Office
9755 Research Drive
Irvine, CA92618, U.S.A.
Phone: 1 (949) 250-4811
Fax: 1 (949) 250-0924
• HORIBA UK Ltd.
2 Dalston Gardens, Stanmore,
MiddxHA71BQ GREAT
BRITAIN UK
Phone: 44 (20) 8204 8142
Fax: 44 (20) 8204 6142
\ts60f/i
m
Printed in Japan TM-T(SK)53
HORIBA
Scientific
NEW
Sulfur-in-Oil Analyzer