GammaScreen-80 pages
ORTEC®
GammaScreen-8™
8-Channel Rapid Analysis Screening of
Multiple Gamma Emitting Radionuclides in Clinical Samples
Easy to Use Rapid and Simple Gamma Radionuclide Screening
• A very compact solution for rapid screening of hundreds of samples/day for possible radioactive
contamination.
• No sample processing: Pass/fail evaluations against analysis limits.
• 700 samples may be counted in 20 hrs of operation.
• MDA (LOD)1 = 52 Bq/L, 300 second count.
• 8 samples (10 ml or 50 ml) counted simultaneously with automated data entry by bar code
scanning (optional).
• 8 completely independent NaI(Tl) well detectors and count chains.
• Digitally stable electronics.
• Full spectral data retained for detailed analysis.
1Minimum Detectable Activity or Limit of Detection.
ORTEC Gross Gamma Counter Sample Evaluation
Counting System #1
Sample Date/Time: Friday, October 16 2009 11:13
Actual (CPS) Limit (CPS) Evaluation
SampleID: 09-911-000064-1GG
Detector: 1 6.36 4.05 FAILED
SampleID: 09-911-000065-1GG
Detector: 2 3.87 3.98 PASSED
SampleID: 09-911-000066-1GG
Detector: 3 3.43 3.80 PASSED
SampleID: 09-911-000067-1GG
Detector: 4 3.54 3.84 PASSED
SampleID: 09-911-000068-1GG
Detector: 5 6.44 3.82 FAILED
SampleID: 09-911-000069-1GG
Detector: 6 3.43 3.68 PASSED
SampleID: 09-911-000070-1GG
Detector: 7 3.55 11.59 Passed
SampleID: 09-911-000071-1GG
Detector: 8 3.58 11.00 Passed