model 90700 pages
TELEDYNE
ANALYTICAL INSTRUMENTS
HEAD SPACE REQUIREMENTS
When testing food packages, the operator pushes a hypodermic
sample tube through a self-adhesive septarium - the instrument
does the rest. A piercing tool is provided for metal containers and
caps. For accurate readings, only 4cc of head space volume is
required.
PRINTER PORT
To print a report, connect a standard printer or personal
computer to the 9070 for a complete summary of every test.
Alternatively, a personal computer or analog recorder can be
connected to the 4-20mA outputs.
MODEL 9070
O2 and CO2 Food
Package Analyzer
• Oxygen and CO2 analysist
• Simple and reliablet
• Low maintenance
SIMPLE TO USE
Teledyne’s Model 9070 Food Package Analyzer is an
ideal unit for spot checking gas measurements within
food containers for ongoing validation of packaging
machines.
After inserting the sample probe into a package or
container, the 9070 automatically draws a sample
and displays the lowest reading of oxygen and
the highest reading of carbon dioxide. There is no
need to touch a key or button during the analytical
process.
ACCURACY WITHOUT CALIBRATION CHECKS
The 9070 uses a robust zirconia oxygen sensor
which provides years of accurate, drift-free
readings from 1 ppm to 100% with only an occasional
reading in air to assure a precise analysis.
The carbon dioxide sensor offers exceptional long
term stability and automatically zeroes itself for every
reading. The sensor only requires calibration once
per week with the recalibration process being a
simple keyboard entry. No other maintenance of the
sensor or unit is required.
SELF DIAGNOSTICS
The programmable alarm relay can be set at any oxygen or
carbon dioxide level or to alert the operator for a variety of other
functions. The advanced alarm system continuously monitors
the sensors’ performance. If a fault is detected, the instrument
immediately alerts the operator and displays the fault.
MODEL DESIGNATIONS
t Modelt
9
t 070-1t
9
t 070-2t
9
t 070-3t
Measurementt
Oxygent
Oxygent
Oxygen & CO2t
Pump
Yest
Not
Yes
ADVANTAGES
• Can be used for spot checks or continuous monitoring
• Long-life, maintenance free sensors
• Only 4cc of head space required for a sample
• Programmable alarm relays
• Light weight and portable
STANDARD FEATURES
• Two isolated, linearized 4-20 mADC outputs
• Computer / printer interface for data
• Self diagnostic alarm
• LCD display content options
• One user-programmable alarm relay
• Accessory kit for various applications