TOC-41100 pages
Specification Sheet
Process analyser
TOC-4110
Measurement type
NPOC (acidify/sparge removal of IC) and TC (standard).
NPOC, TOC (TC-IC) (option).
NPOC,TOC(TC-IC and POC+NPOC)(option)
Measurement principle
Catalytically aided combustion with NDIR detection
Measurement range
Variable from 0-5ppm full scale to 0-1000ppm full scale
(0-20,000ppm is possible with on-board dilution function)
Reproducibility
Within +/-2% of full scale
Span stability
Within +/-2% of full scale/day
(with temperature fluctuation within 5° C)
Zero stability
Within +/-2% of full scale/week
Measurement cycle
Minimum 4 minutes (NPOC)
Sample dilution
Dilution with syringe, variable sample dilution factor of 2-20
Auto calibration
1-2 point auto calibration possible
Multi-stream option
2 streams with optional sample switching set,
Maximum 6 streams with optional flow line switcher
Display
LCD 40 characters x 14 lines
Recorder option
Recorder width is 100mm, 6 dot recorder
Printer option
Thermal type, 40 columns, chart width with 110 mm.
Analog output
Selectable output of 0-1VDC, 0-16mA or 4-20mA
(insulated output)
Output
RS 422
Alarm
Selectable among upper/lower limit, upper/upper limit and lower/lower limit and instrument trouble alarm
Input signal
Sample measurement start, sample measurement stop, calibration start, alarm reset
Event signals
Ready state, operating online, measuring, calibrating and more.
Carrier gas
Air or oxygen, supply pressure:300kPa, free of dust, oil mist, water droplets
Power
100,110,115,120,220,230,240V as ordered; 5A;50-60Hz
Construction
Indoor wall mounting (optional stand available)
Shimadzu Europa GmbH www.shimadzu.de
Albert-Hahn-Str.6-10, 47269 Duisburg