LC-2010 Integrated HPLC System0 pages
Shimadzu Europa GmbH www.shimadzu.de
Albert-Hahn-Str.6-10
47269 Duisburg
Specification Sheet
LC-2010
Integrated HPLC System
LC-2010C LC-2010A
Pump type Serial dual plunger, micro volume
(10 ìL on primary side,
5 ìL on secondary)
Flow rate setting range 0.001 ~ 5 mL/min.
(1.0 - 35.0 MPa)
Flow rate accuracy ± 1% (1 mL/min, 3.0 - 10.0 MPa,
using water and fixed ambient
temperature of 20°C, except
during gradient solvent delivery),
± 2% (except for above
conditions) or ± 2 ìL/min,
whichever is greater
Flow rate precision 0.075% RSD or 0.02 min SD,
whichever is greater
(0.05 - 2 mL/min, 1.0 - 35.0 MPa,
methanol 20°C constant
temperature)
Number of solvents
and setting range
Up to 4 solvents
0 to 100%, 0.1% steps
Concentration
accuracy
± 0.5% (aqueous acetone/water,
binary gradient, flow rate 0.1 - 2.0
mL/min, pressure 1.0 - 20.0 MPa)
Pump
LPGE
Concentration
precision
± 0.1% (aqueous acetone/water,
binary gradient, flow rate 0.1 - 2.0
mL/min, pressure 1.0 - 20.0 MPa)