Strip Chart Recorder ES6000 pages
ELECTRONIC MINI-RECORDER
CHINO
PERFORMANCE SPECIFICATION
MODEL ES6ED
The ES600 is a compact industrial analog recorder using
60-mm wide chart paper. Even with its compact size of
96 x 144 mm, it has functions comparable with a large-
size recorder including a highly reliable automatic parallel
system, easy-to-install fanfold recording chart paper, etc.
There are two recording systems, pen type recorders have
one-point continuous recording and dot printing type
recorders are available which record a maximum of six
points. Thermocouples, DC voltages/currents, resistance
thermometers and thermistors measuring inputs are avail-
able which enable the recording of temperatures, various
industrial units, signal voltages and currents.
■ GENERAL SPECIFICATIONS
MODEL ES600
Input signal
Scale width
Indicating accuracy
Dead band
Balancing speed
Chart paper
Recording points
Recording system
Chart speed
Dot printing Interval
Pen lit!
(Pen-wriling type)
Power supply
Allowable voltage
fluctuations
Ambient temperature
Ambient humidity
Allowable signal
source resistance
Input resistance
Max. common mode
voltage
Common mode
rejection ratio
mV: More titan 10 mV DC span,
less than 50 V span
Thermocouple: K, T More than 250°C span
E, J More than 200°C span
R More than 1000°C span
Resistance thermometer: (PtlOO, JPT100)
More than 50°C span
Thermistor: More than 50°C span
(at room temperature)
60 mm
±1.0% of input span
0.4% of input span
Approx. 2.0 sec (50 Hz), 1.6 sec (60 Hz) for
full-scale movement
Fanfold type, effective width 60 mm,
total width 73 mm, total length 10 m
Pen type; 1 pen
Dot printing type; 4 types with 1, 2. 3 and
6 points
Pen type: Cartridge pen continuous
recording (color: red)
Dot printing type: Ink pad dot printing will)
each point in a different
color
Dot printing color:
1 point: red
2 point: 1 red, 2 blue
3 points: 1 red, 2 blue, 3 green
6 points: 1 red, 2 blue, 3 green, 4 violet,
5 purple, 6 brown
10,20 mm/h and last feed
Appro*. 10 sec
Manual
100, lit), 120, 130.200, 220 230,
or 240 V AC, 50/60 Hz
+10% to -10% of rated value
10°C to 5()°C
30 to 90% RH
mV input, thermocouple input: 1 Kli or less
Resistance thermometer input: less than
10 Q per wire
Approx. 8 Mil
200 V AC
100 dB or more
Series mode rejection
ratio
Insulation resistance
Withstand voltage
Power consumption
50 dB or more
Between measuring terminals and ground
terminal: 500 V DC. 20 Mil or more
Between power terminals and ground
terminal: 500 V DC, 20 Mil or more
Between measuring terminals and power
terminals; 500 V DC, 20 Mii or more
Between measuring terminals and ground
terminal: 500 V AC for 1 minute
Between power terminals and ground
terminal:1000 V AC for 1 minute
Bel ween measuring terminals and power
terminals: 1000 V AC for 1 minute
Approx. 7 VA
Casing
Color
Mounting
Weight
Front door: ABS plastic
Flange and rear case: ABS plastic
Gray
Panel mourning
Approx. 1.7 kg
MODELS
□
I
- Input
-Alarm
- Number 01
of points ID
02
03
06
Thermocouple
Resistance thermometer
Thermistor
Cunent or voltage
None
High/Low limit*'
High limit*
Low limit*
1 pen
1 dot
2 dots
3 dots
6 dots
CHINO
Functions marked * are optional.
DCE.1RO
"