HBM-100A Panel Type pH Analyzer/Controller0 pages
Panel Type pH Analyzer/Controller
Panel Type ORP Analyzer/Controller
HBM-100A
HBM-102A
Compact DIN size (96 x 96 mm) panel mount pH/ORP
controllers.
2-point alarm contact output and 4~20mA DC
transmission output are equipped as standard.
The controller is equipped with an automatic, single-action
stability judgment function, which allows for accurate
calibration using standard solutions and helps to eliminate
operator errors. During calibration, the controller determines
the status of the electrode by monitoring its characteristics
and displays diagnostic information in the form of
messages.
Output (measurement) range that can be freely set.
Measurement and display of sample temperature
(0~100ºC).
When the controller enters maintenance (ST-BY) mode, the
previous output value is held and the alarm output is
disabled. This helps to prevent disruptions to the control
system, such as chemical feeds.
The controller automatically switches back to measurement
mode if it is left in maintenance (ST-BY) mode for more
than two hours.
Crack detection function for pH electrode.
Temperature compensation for sample pH value and
pH/ORP value shift functions.
Standard Specifications
Product name
Model
Measurement range
Panel type pH analyzer/controller
Panel type ORP analyzer/controller
HBM-100A
HBM-102A
mV: -600 ~ +600 mV
pH: -1.00 ~ 15.00
mV: -2000 ~ +2000 mV
Temp: 0 ~ 100ºC
Temp: 0 ~ 100ºC
Display only. No output-signal.
Display only. No output-signal.
Transmission output signal
4~20 mA DC, isolated. Max. resistance: 650Ω or less.
Transmission output range
Adjustable (0.1 pH steps). Minimum width of 2 pH.
Adjustable (10 mV steps).Minimum width of 400 mV.
Control (alarm) contact
2 contacts (upper and lower limits can be set freely) a-contacts Contact capacity: 250V AC, 3A or less (resistive load)
output
Control sensitivity: Set to a value between 0.01~2.00pH
Control sensitivity: Set to a value between 1~200mV
Linearity: ±0.03pH or less (using equivalent input)
Linearity: ±3mV or less (using equivalent input)
Performance
Repeatability: ±0.02pH or less (using equivalent input)
Repeatability: ±3mV or less (using equivalent input)
Response: 5 sec. or less for 90% response in FA (fast) mode, 50 sec. or less for 90% response in SL (slow) mode.
Power requirements/Power consumption
90~264V AC, 50/60Hz approx. 5VA
Ambient conditions
-10~50ºC, 0~90% RH (when in transport and storage: -30~70ºC, 0~99% RH)
Dimensions/Weight
96 (W) x 96 (H) x 90 (D) mm (panel cut-out 92 x 92 mm), approx. 0.5kg
Cleaning signal input: The controller can receive a “cleaning” signal from the chemical and brush cleaners to hold output during the cleaning process.
Temperature compensation for sample pH value: Coefficient setting range…±0.100/ºC Standard conversion temperature…25ºC
Other functions
Manual temperature compensation for glass electrode: Manual temperature compensation is carried out by specifying the sample water temperature.
pH/ORP value shift: Measured value can be shifted within the range of ±1.00 pH/±100 mV. (Temperature shift range: ±5ºC)
Burnout: Output signal can be shifted to the upper or lower limit when there is an abnormality, such as a cracked glass membrane or temperature sensor failure.
Alarm contact output: 2 extra contacts (a-contacts, upper/lower limits can be set freely, no bandwidth limit) (4 contacts in total)
Optional features
Cleaner control output: The internal timer delivers 100V AC power to the chemical cleaner, brush cleaner, and other cleaners.
RS-232 output: A digital communication cable can be used to send the measured pH/ORP and temperature values to a computer.
Configuration
Detector
HBM-100A/102A
Electrode lead
EC-10 Extension cable
5m
(Max. 100m)
Cleaner
Transmission output
DC 4
20mA
Receiver
Junction box
FC-4
Cleaning signal
To cleaner power supply (100V AC), option
ON-OFF
AC 90~264V
Alarm contact output
Power supply
Recorder
Controller, etc