Autosamplers0 pages
www.discoverysciences.com 15
hplc instruments
analytical hplc
Alltech® 580 Autosampler
• Automates sample
prep tasks: sample
dilution, derivatization,
and internal standard
addition
• Capable of three
different injection
techniques
• Large injection volume
range for maximum
versatility
• Built-in column heater
Autosamplers
Alltech® 570 Autosampler
• Affordably
automate HPLC
and IC analysis
• Flushed loop
filling maximizes
precision and
reproducibility
• Partial loop filling
minimizes sample
waste
1408
Spare Parts for 570 Autosampler
Description Part No.
Pump Tubing Cassette 575115
Side-Port Sample Needle, 3/pk 575130
Large Volume Option, Field Installable 575165
Air Needle 575125
Sample Drain for Headspace Pressure 575170
Auxilliary and Control I/O Cable 570101
Tubing from Needle to Valve, 10ft 20064
Needle Connection Nut 575175
Needle Connection Ferrule, 10/pk 575185
Standard Vial Tray 575110
Model 580 Autosampler
Description Part No.
Model 580 Autosampler
Standard Tray, SS Valve 580100
Standard Tray, PEEK Valve 580500
Tray Options
Standard Volume Tray, 84 Vial Capacity, 1.5mL Vials
with 3-10mL Vial Positions
580201
Large Volume Tray, 24 Vial Capacity, 10mL Vials 580200
Large Capacity Tray, 96 Vial Capacity, 1.5mL Vial 580265
Model 570 Autosampler
Description Part No.
Flushed Loop Filling
Stainless Steel 575100
PEEK 575500
Partial Loop Filling
Stainless Steel 570150
PEEK 570550
570 Autosampler Specifications
Vial Capacity: 96 samples (92 with needle wash), 48 (5mL)
vials (optional)
Injection Volume: Flushed Loop: 5–500ìL (dispenser controlled),
5–5000ìL (by head space pressure).
Partial Loopfill: 10–490ìL in steps of 10ìL
(depending on the loop volume, maximum 50%
of the loop volume)
Injections per Vial: 9 Maximum
Needle Wash: Selectable, 50ìL fixed volume between vials
Reproducibility: Flushed Loop Injections: RSD0.5%
Partial Loop Injections: RSD1.0%
Power: 115 VAC/230 VAC; 50/60 Hz
Dimensions: 11.1" H x 11.8" W x 17.3" D
(28cm H x 30cm W x 44cm D)
Weight: 34lb (15.5kg)
Warranty: 1 year
580 Autosampler Specifications
Vial Capacity: 84 (1.5mL) vials + 3 (10mL) vials,
96 (1.5mL) vials (optional),
24 (10mL) vials (optional)
Injection Volume: 1–5000ìL in 1ìL increments,
5000–7500ìL in 1.5ìL increments,
7500–10,000ìL in 4ìL increments
Injections per Vial: 9 maximum
Analysis Time: 9hrs 59min 59sec maximum
Reproducibility: Full Loop Injections: 0.3% RSD,
Partial Loop Injections: 0.5% RSD,
ìL Pick-Up Injections(>10ƒÊL):1.0%
Carry Over: <0.05% with needle wash
Column Oven Temp.
Range:
Ambient +5°– 60°C
Temperature
Controlled Zone:
11.8" H x 1" W x 1" D
(30cm H x 2.5cm W x 2.5cm D)
Power Requirements: 115 VAC/230 VAC; 50/60Hz
Dimensions: 13.4" H x 11.8" W x 19.7" D
(34cm H x 30cm W x 50cm D)
Weight: 33lb (15kg),
Warranty: 1 year
Multifunctional autosampler
improves laboratory efficiency.
Choose between flushed and partial
loop filling and stainless steel and PEEK
fluid paths.
3000