Accuracy Thermometer PHP 6020 pages
نسخه متنی
"
"
Main applications include: Temperature measurements using RTDs Օ Absolute or differential measurements over two channels. Calibration of temperature sensors. Օ Differential thermal analysis. Checking the temperature stability of furnaces or baths. Օ Monitoring (alarm) and temperature recording. | ||||||||||||||||
i Resolution 0.0001°C 1 Accuracy 0.009°C RTDs Dual inputs Օ RS 232 and IEEE 488 interfaces | ||||||||||||||||
The PHP 602 is a high accuracy, dual input thermometer suitable for use with RTDs. If a large number of probes already exists in the data base, the user can at anytime, Resistance measurement (1)................ | create new linearisation tables and the metrological parameters (correction versus standards, serial numbers, names, calibration dates ...) are stored in the memory. | Thanks to these functionalities, PHP 602 is perfectly designed for accurate probes calibration using comparison method and studies of temperature phenomena. | ||||||||||||||
All ranges 25 to 3 200 n | Stability 24 hours 0.0005%Cal | Accuracy 90 days (2) 0.0030°% + 0.0005°% | ||||||||||||||
Accuracy 1 year (2) 0.0045°% + 0.0005°% | ||||||||||||||||
(1) 4 wire configuration, all currents. For the 3 wire configuration, add ImCl. (2) Accuracy is expressed in ± (% reading + % range) at 23°C ± 1°C. | ||||||||||||||||
Three current waves: direct, pulse and alternate, together with I/V2 function to define self-heating measurement. 3 or 4 wire configuration. Measurement of all types of sensors as described below: - Standardized sensors according to IEC Publication 751/1995, a = 3851, Pt 100, 200, 500 and 1000. JISC 1604/1989, a = 3916, JPt 100. EIT 90, a = 3926, Pt 100. DIN 43760, a = 618, Ni 100. MIL-T 24388C, a = 672, Ni 120. __ | MINCO 16/9, a = 427, Cu 10. - Callendar and Van Dusen equations defined by coefficients (Ro, A, B, C) or by 4 couples of points (resistance/temperaͭture). - EIT 90 equations defined by R at 0.01°C and the deviation function coefficients or by resistance values at fixed points. - Polynomial equations defined per point (25 points max.). | |||||||||||||||
Temperature measurement_________________ Direct reading in °C, °F or K. - From 1 to 4 calibration points may be entered for all types of sensors. - Digital filter. - Programming using either the RS 232 link (standard) or the IEEE 488 link (option). - Storage of up to 5000 measurements together with date. - Measurements triggered by internal or external event. - 2 alarm outputs relays. - Choice of 6 measurement current from 0.125 to 4 mA. |