22000 pages
SERVOTOUGH OxyExact (2200)
nnnnPROCESS OXYGEN ANALYSERS
nnnnfor use on demanding process control applications
nnnnincluding safety critical, purity, feedstock clean-up
nnnnmeasurements where high performance is critical
nnnnUnrivalled accuracy, reliability and repeatability
nnnnLow cost of long term ownership
nnnnUp to six transmitters can be connected to a
nnnnsingle control unit
nnnnHazardous area ATEX Zone 1, FM, CSA, TlIS, lECEx
nnnnTransmitter interface software available
nnnnEthernet or RS485 digital communication options
nnnnusing Modbus
nnnnSPECIFICATION
nnnnTRANSMITTER
nnnnTRANSMITTER
nnnnCONTROL UNIT
nnnnCONTROL UNIT
nnnnTECHNOLOGY
nnnnSample Gas Temperature
nnnnHazardous area:
nnnnPERFORMANCE
nnnnMaximum range:
nnnnMinimum range:
nnnnIntrinsic Error (accuracy):
nnnnLinearity:
nnnnRepeatability:
nnnnResponse time (T90):
nnnnZero drift per week:
nnnnSpan drift per week:
nnnnParamagnetic
nnnn60°C *
nnnnATEX Cat. 2
nnnnlECEx
nnnnFM Class I Div. 1
nnnnCSA
nnnnTIIS
nnnn0-100%
nnnn0-0.5%
nnnn<0.02% 02
nnnn<0.01% 02
nnnn0.02%
nnnn<4 seconds
nnnn<0.02% 02/week
nnnn<0.05% 02/week
nnnnParamagnetic
nnnn<110°C *
nnnnATEX Cat. 2
nnnnlECEx
nnnnFM Class I Div. 1
nnnnCSA
nnnn0-100%
nnnn0-1%
nnnn<0.04% 02
nnnn<0.02% 02
nnnn0.03%
nnnn<4 seconds
nnnn<0.08% 02/week
nnnn<0.10% 02/week
nnnnATEX Cat. 2
nnnnlECEx
nnnnFM Class I Div.1
nnnnCSA
nnnnTIIS
nnnnATEX Cat.3
nnnnlECEx
nnnnFM Class I Div.2
nnnnCSA
nnnnTemperature of operation: Consult Servomex for other temperatures
nnnnSERVOMEXX
"