T701H0 pages
TELEDYNE
ADVANCED POLLUTION INSTRUMENTATION
A Teledyne Technologies Company
T SERIES
Model T701H High Performance Zero Air System
The T701H adds Modbus TCP/IP connectivity to the proven
reliability and performance of our high purity zero air system. This
allows remote monitoring of key instrument functions including
dewpoint status, scrubber temperature, box temperature and
output pressure. Additionally, this digital access provides the
ability to remotely disable the internal pump, allowing longer
service intervals and less power consumption. The user can also
remotely start instrument warm-up prior to arrival at the site.
TheT701H is a fully self-contained source of high purity zero air
for dilution calibrators. It is ideal for use with highly sensitive
analyzers in ambient background and trace level applications.
TheT701H includes a dewpoint sensor, an oil and diaphragm-free
pump, scrubbers to remove SO2, NO, NO2, O3 and H2S, plus
high performance scrubbers to remove CO and Hydrocarbons,
►► Modbus Connectivity
►► Trace Level output purity
►► Dewpoint Sensor
►► Once/year maintenance interval
►► Regenerative, heatless dryer for
maintenance-free water removal
►► Long-life scrubbers for SO2, NO,
N02, O3 and H2S
►► High performance CO/Hydrocarbon
scrubber
►► Automatic water drain
►► Automatic pump control with demand
tank
►► Two-year warranty
The regenerative, heatless dryer removes water and produces
output with a dew point less than -40°C (up to 30 SLPM flow
rate). The system's pressure, temperature, and dewpoint values
are all continuously monitored, which allows a microcontroller to
adjust the pump cycling frequency, valve timing, and heater
power for optimal performance.
Free Custom&r Support by t&l&phon& and email for the life of the instrument
"