A3000 pages
Amino Acid Analysis
Technical Specifications
Pump
Reactor
Simultaneous 2- piston technology of reagent and eluent lines
Pump head made of titanium material
Max. pressure: 400 bar
Flow rate: 0,001 - 9,999 mL/min
Reproducibility of flow rate: 0,1 % (RSD) at 100 µL/min
Highly inert material
Peltier elements
Temperature range 50 - 150° C
Auto-Shut-down safety routine in case of failure alarm
Tubings
32 bit software iControl
including optimized separation programs
comfortable creation & modification of separation programs
automatic log-file generation of analysis parameters
PEEK and FEP
Control- Software
A300
Amino Acid Analyzer
Microphotometer
Wave length: 570 nm, 440 nm
Noise: 0,02 mAU, 1s
Drift: < 1,0 x 10 -5 AU / h
Linearity deviation: 0,3 % bei 1 AU
Measuring range: high, low
Analog signal: 0 - 5 V
Base line adjustment: free programmable
Auto-Zero function: free programmable
Flow cell: 10 µL volume
maintenance free
Data-acquisition/ handling Software - Options
Data- acquisition and Data- analysis by aminoPeak software
specially designed for amino acid analysis,
fast and user-friendly
21 CFR Part 11 conform Data- acquisition and Data- analysis by Clarity
software to work in regulated environments
Interfaces
Separation column
Cation exchange resin: 3μm or 7µm
Stainless steel
Temperature adjustment by Peltier elements
Temperature range: 20 - 100° C
Temperature accuracy: 0,1° C
2 USB Ports for A300- operation and
signal registration of photometers
Ready to use reagent kits
Kits including eluents, reagent and additonal dilutions buffers, separation
and ammonia absorber column for 500 or
1000 analytical runs
Autosampler
sample cooling as low as 4 O C possible by peltier elements
Syringe pump, motor- injection valve
Sample rack: 2 x 48 Vials (1,5 mL Standard), 4 x 48 as option
Dosage: full loop (100 µL) or 1 - 25 µL in 1 µL- steps free
programmable
Reproducibility of retention time: <0,1% (Arg), RSD
Reproducibility of peak area:t0,8%, CV 0,5% (Gly), CV 0,6% (His), RSD
Detection limit: 3pmol (Asp), 1pmol (His)
Total retention time: 30 min. for 18 amino acids, 45 min. for 20 amino acids
Eluents and Reagent
Reagent- and washing solution, in central unit
5 eluent bottles, 1 regeneration bottle in eluent rack
All information is subject to change without further notice
For further information please contact:
tttt
membraPure GmbH t
Neuendorfstraße 20at
16761 Hennigsdorf/Berlint
Germany tt
ttt
Phone: +49 (0)3302/ 202 35-00 t
Fax:
+49 (0)3302/ 202 35-01t
ttttt
E-mail: info@membrapure.de
Web:
www.membrapure.det
innovative technology combined
with classical method