Dynamometer Controller/Readout 62000 pages
M AGTROL
6200
Data Sheet
Model 6200 Open-Loop
Dynamometer Controller
Features
•t
•t
•t
•t
•t
•t
•t
•t
•t
•t
•t
•t
Open-Loop Dynamometer Control
Built-in Pass/Fail Motor Testing Capability
Interfaces: RS-232 and IEEE-488
High Speed Data Acquisition: 120 torque and speed
points per second via IEEE bus (approx. 60/sec. via
RS-232)
High Quality, Easy-to-Read Vacuum Fluorescent
Readout: Displays torque, speed, power and auxiliary
values
Current-Regulated Supply: Provides up to 1 amp
output
Adjustable Torque Units: English, Metric and SI are
standard
Dynamometer Overload Protection
Internal Data Storage: Up to 100 data points
Auxiliary ± 5 V DC Analog Input: For additional
transducer
Closed Box Calibration
Rack Mounting: 19" (482.6 mm) with handles
Description
Magtrol's Model 6200 is an Open-Loop Controller designed
for use with any Magtrol Hysteresis Dynamometer. The unit
provides open-loop control of the dynamometer via an internal
current-regulated power supply. With a high-quality vacuum
fluorescent readout, the Model 6200 displays torque, speed
and mechanical power values of the motor under test. In place
of mechanical power, it can also display auxiliary transducer
readings via the ± 5 V DC analog input. These displayed
values can be stored internally or output via the RS-232 or
IEEE-488 interface.
Pass/Fail Motor Testing
The Model 6200 comes with an easy-to-use motor testing
Pass/Fail feature. This feature is ideal for quick pass/fail
(go/no go) testing in production and incoming inspection
applications.
When the 6200 is operated in the Pass/Fail mode, one of three
readings is used as the tested parameter: torque, speed or the
auxiliary transducer. The two parameters not used are set with
user-defined upper and lower acceptable limits. As the motor
is loaded to the tested parameter value (for example, speed),
the other two parameters (in this case, torque and transducer)
are measured. Test results (for the other 2 parameters) are
indicated with a “PASS” or “FAIL”, or the display can be
toggled to show the actual values.
System configuration
Hysteresis Dynamometer
Motor
Under
Test
MODEL 6200 CONTROLLER
CAUTION: DOUBLE POLE FUSING
75VA 50/60 Hz
BRAKE
ACCESSORY
TORQUE-SPEED
OUTPUT
CTRL OUT
DYNAMOMETER
AUX.
INPUT
RS–232C
GPIB/IEEE–488
EARTH
GROUND
BRAKE FUSE
(5×20mm): UL/CSA 1.25A 250V SB
IEC 1A 250V T
MAGTROL, INC. BUFFALO, NY
FUSE (5×20mm):
120V UL/CSA 800mA 250V SB
240V IEC 315mA 250V T
Ordering Information
6200t Open Loop Dynamometer Controller 120 VAC
6200At Open Loop Dynamometer Controller 240 VAC
www.magtrol.com