nVision Reference Recorder MPa
Pressure Module (PM)
70 MPa
ACCURACY
MPa (Gauge Pressure)
MPaA (Pressure with BARO module)
300 kPa, 1 MPa, and 3 MPa modules
35.5
(1.40)
300 kPa module
0 to 30% of Range:t ±(0.0075% of Full Scale)
t 30 to 110% of Range:t ±(0.025% of Reading)
tVacuum:t ±(0.06% of Full Scale*, typical)
tt
* Full Scale = -99.9 kPa
10 and 30 MPa modules
t
0 to 30% of Range:t ±(0.015% of Full Scale)
t 30 to 110% of Range:t ±(0.05% of Reading)
70 and 100 MPa modules
t
0 to 30% of Range:t ±(0.03% of Full Scale)
t 30 to 110% of Range:t ±(0.1% of Reading)
Includes all effects of linearity, hysteresis, repeatability, temperature,
and stability for one year.
All models indicate vacuum, but vacuum specification (typical) applies
to 300 kPa, 1 MPa, and 3 MPa models only.
Not recommended for continuous use at high vacuum.
Refer to XP2i-DP data sheet for gauges that are intended for continuous
high vacuum use.
AA
1.38 to 100.00 kPaA:t ±0.08 kPaA, typicalt
t 100.00 to 400.00 kPaA:t ±(0.025% of Reading)+0.03 kPaA
14.7
(0.58)
1 MPa module
0.00138 to 0.10000 MPaA:t ±0.00008 MPaA, typical
t
0
t .10000 to 0.40000 MPaA:t ± 0.00010 MPaA
0
t .40000 to 1.10000 MPaA:t ±(0.025% of Reading)
3 MPa module
t 0.0014 to 0.1000 MPaA:t ±0.0001 MPaA, typical
t 0.1000 to 1.000 MPaA:t ± 0.0003 MPaAt
t 1.000 to 3.1000 MPaA:t ±(0.025% of Reading)
10 MPa module
View AA
t 0.1000 to 3.1000 MPaA:t ± 0.0015 MPaA
3
t .1000 to 10.1000 MPaA:t ±(0.05% of Reading)
30 MPa module
t
0.100 to 9.100 MPaA:t ± 0.005 MPaA
t 9.100 to 30.100 MPaA:t ±(0.05% of Reading)
70 MPa module
t 0.100 to 21.100 MPaA:t ± 0.021 MPaA
t 21.100 to 70.100 MPaA:t ±(0.1% of Reading)
100 MPa module
t 0.100 to 30.100 MPaA:t ± 0.030 MPaA
3
t 0.100 to 100.100 MPaA:t ±(0.1% of Reading)