nVision Reference Recorder bar Datasheet0 pages
nVision Reference Recorder bar
Pressure Module (PM)
700 bar
ACCURACY
bar (Gauge Pressure)
barA (Pressure with BARO module)
3, 10, and 30 bar modules
35.5
(1.40)
3 bar module
0 to 30% of Range:t ±(0.0075% of Full Scale)
t 30 to 110% of Range:t ±(0.025% of Reading)
tVacuum:t ±(0.06% of Full Scale*, typical)
tt
* Full Scale = -1.0 bar
100 and 300 bar modules
t
0 to 30% of Range:t ±(0.015% of Full Scale)
t 30 to 110% of Range:t ±(0.05% of Reading)
700 and 1000 bar modules
t
0 to 30% of Range:t ±(0.03% of Full Scale)
t 30 to 110% of Range:t ±(0.1% of Reading)
Includes all effects of linearity, hysteresis, repeatability, temperature,
and stability for one year.
All models indicate vacuum, but vacuum specification (typical) applies
to 3, 10, and 30 bar models only.
Not recommended for continuous use at high vacuum.
Refer to XP2i-DP data sheet for gauges that are intended for continuous
high vacuum use.
AA
t0.0138 to 1.0000 barA:t ±0.0008 barA, typicalt
t1.0000 to 4.0000 barA:t ±(0.025% of Reading)+0.0003 barA
14.7
(0.58)
10 bar module
t0.0138 to 1.0000 barA:t ±0.0008 barA, typical
t1.0000 to 4.0000 barA:t ± 0.0010 barA
4
t .0000 to 11.0000 barA:t±(0.025% of Reading)
30 bar module
t 0.014 to 1.000 barA:t ±0.001 barA, typical
t 1.000 to 10.000 barA:t ± 0.003 barAt
t10.000 to 31.000 barA:t ±(0.025% of Reading)
100 bar module
View AA
t 1.000 to 31.000 barA:t ± 0.015 barA
t31.000 to 101.000 barA:t ±(0.05% of Reading)
300 bar module
t
1.00 to 91.00 barA:t ± 0.05 barA
t 91.00 to 301.00 barA:t ±(0.05% of Reading)
700 bar module
t 1.00 to 211.00 barA:t ± 0.21 barA
t211.00 to 701.00 barA:t ±(0.1% of Reading)
1000 bar module
t 1.00 to 301.00 barA:t ± 0.30 barA
t301.00 to 1001.00 barA:t ±(0.1% of Reading)
4792. G nVision bar Data Sheet
Page 1 of 10