LR-Cal QUARTZ-350 pages
Portable Temperature Calibrator QUARTZ
Dry Block Calibrator -50°C...+150°C
QUARTZ-35
Dry Block Temperature Calibrator QUARTZ-35
Temperature range: -30°C up to +150°C (at 20°C ambient temperature)
Type QUARTZ-35-2I: with 2 inputs for ext. reference and test specimen
C for mobile or stationary use
C micro processor operated (PID)
C ramp function, thermostat test
Typical applications:
C calibration and checking of all kind of thermometer
C calibration and checking of thermostats
C computer aided calibration via RS 232 interface
Technical Data:
Temperature range: -30°C...+150°C (at 20°C ambient temperature)
Accuracy of temperature reading: ±0.15°C
Integrated reference probe: Resistance thermometer (3-wire), class A
Display resolution: 0.01 / 0.1°C (display switchable to °F and K)
Stability of regulated temperature: ±0.03°C (at 0°C)
Mean heating time: 20°C per minute / Mean cooling time: 22°C per minute
Uniformity (at 0°C): radial ±0.02°C, axial ±0.1°C
Ventilation: 2 vents, electronic regulated
RS232 interface: included (build-in)
Ramp function: min. 0.1°C per minute
Dry block: 140 mm dept, 35 mm diameter
Usable dept: 130 mm
Power Supply: 230 VAC (optional: 115 VAC)
Power consumption: 300 VA
Case: metall, dimension 160 x 340 x 330 mm
Weight: 10 kg net (shipment weight: 17 kg)
NOTE: The stated accuracy is valid for ambient temperature +10°C...+40°C and is guaranteed for 1 year.
Scope of standard Supply:
Temperature Calibrator QUARTZ-35, power supply cable, fuses kit, thermostats connection cables, operating manual, certificate of
calibration (traceable), tweezers for inserts removing.
1 insert with 6 holes (4 - 4.5 - 5.5 - 6.5 - 8.5 - 10.5 mm)
Additional for type QUARTZ-35-2I:
Version with data acquisition card and two input devices to measure Pt 100 (3-/4-wire, -100...+700°C, ±0.3°C), Thermocouple type J
(0...1000°C, ±1°C) and Thermocouple types K/N/R/S (0...1300°C, ±1°C).
Optional Accessories:
• Insert without holes (to be drilled by customer) • Insert with 3 holes (4.5 - 6.5 - 9.5 mm diam.)
• Insert with 2 holes (6.5 - 19.5 mm)
• Insert with 1-3 holes as per customers request, diameter 3.5 to 20 mm
• Insert with 4-6 holes as per customers request, diameter 3.5 to 20 mm
• Soft bag
• Alu case
C PC-Windows Software “AQ2sp” and RS232-Kabel
With the software “AQ2sp” it is possible to remote-control the calibrator
with a PC. Automatic or manual calibration of one or more test specimen,
stress tests, life cycle tests, thermostat tests. Printing of calibration
certificates.
DRUCK & TEMPERATUR Leitenberger GmbH
Bahnhofstr. 33 C D-72138 Kirchentellinsfurt C Germany
Tel.: +49 - 71 21 - 9 09 20 - 0 C Fax: +49 - 71 21 - 9 09 20 - 99
E-Mail: DT-Export@Leitenberger.de C http://www.druck-temperatur.de
Technische Änderungen vorbehalten. Freibleibend C (Rel. 100121) • All technical modifications reserved. Without engagement