AMC3100 Series0 pages
ALIAMAG
®
Converter For Electromagnetic Flowmeter
Model AMC3100 Series
GENERAL
ALIAMAG The Aliamag AMC3100 series converter is a microprocessor
based intelligent transmitter that can be paired with any of the Alia AMF
series sensors. The unit features multiple display parameters including
flow rate, flow total, diagnostics, programming and troubleshooting.
FEATURES
Three line LCD display Flow Rate / Totalizer
Accuracy of +/-0.2% value of reading
Bi-directional flow measurement, current and pulse output
Explosion Proof, Exd(ib)ibqllBT5
Records up to 99 power on / off time
MODBUS/HART communication
STANDARD SPECIFICATION
Excitation
: DC pulsed
Display
: 3 lines LCD with illumination
Keypad
and display control
Line 1 : 11 digit totalizer for forward flow
Remote switch instead, no need to open
: 11 digit totalizer for reverse flow
: 11 digit totalizer for difference
: 3 internal keys for programming
cover
Low Flow Cutoff
: 0 to 9.9 %
Line 2 : 5 digit flowrate / velocity / % / mA / HZ
Data Logger Function
: 99 times flowrate records for power on/off
Line 3 : Parameter / Query / Clr tot / Query LOG
Record Data
: Date/Time when Power on/off
Alarm, Guidance
: Flow rate when Power on/off
Unit : M3/H, L/H, Kg/H, T/H, L/min, M3/min, GPH
: Totalizer when Power on/off
: GPM, ML/H, ML/m, ML/sec
Damping
: 0 to 99 Seconds
: Min. 0 to 0.15 m/s
Density Setting
: 0.1 to 9.99 g/cm3
: Max. 0 to 15 m/s
Measuring Range
Self Diagnosis
: The following error message are
: +/-0.2% of reading (Velocity ≥ 0.5 m/s)
displayed if failure occurs
: +/-0.001 m/s (Velocity < 0.5 m/s)
- Coil failure
Repeatability
: +/-0.05% of reading
- Power failure
Current Output
: 4-20 mA (Isolated)
- Overanged Output
Accuracy
Load : Max. 600 Ω
Pulse Output
- Internal error
: Open collector
- Overflow
Rating : 3 to 30 Vdc, 20 mA Max.
Pulse Rate
- Empty pipe
: 1)Scale pulse (Pusle/M3, Pulse/L, Other)
Cable Entry
: Standard :M20 / PG11 Option:1/2"NPTF
: 2)Frequency: 0-5000 Hz
Protection Class
: IP 67, Explosion proof, Exd(ib)ibqllBT5
Pulse Width
: 50% of band width
Ambient Temperature : -25 to +65 °C
Alarm
: High/Low with open collector output
Material
: Aluminum Alloy
Power Supply
: 90-260 Vac, 50/60 Hz
: HART signal (Compatible)
Power Consumption
: 3-17 W (Dependent on sensor size)
: Operation parameters and totalizer
Weight
: 4.5 Kg
Mounting
: Wall mounting
Rating : 3 to 30 Vdc, 20 mA Max.
Communication
Data Storage
: RS 485 (MODBUS Protocol)
figures are stored by EEPROM for more
than 10 years
ALIA GROUP INC.
113 Barksdale Professional Center, Newark, DE 19711, USA
TEL : + 1 - 302 - 213 - 0106
FAX : + 1 - 302 - 213 - 0107
: 24Vdc +/-10% (Option)
: Bracket on 2" pipe
URL : www.alia-inc.com
E-mail : alia@alia-inc.com
AMC3100V1.1.9.A4.en