XP2i Digital Pressure Gauge psi Data Sheet0 pages
XP2i Pressure Gauge psi
ACCURACY
0 to 20% of Range:t ±(0.02% of Full Scale)
t 20% to 110% of Range:t ±(0.1% of Reading)
tVacuum:t ±(0.25% of Full Scale*)
tt
* Full Scale = -14.5 psig
OPERATING TEMPERATURE
Temperature Range:t -10 to 50° C (14 to 122° F)
DISPLAY
tScreen:t 5.5 digits
t
Single Line Display:t16.9 (0.67") high
mm
t
Dual Line Display:t Main Display: 13.9 mm (0.55")
ttSecondary Display: 5.3 mm (0.21")
t
Display Rate:t 4 readings/second (standard)
tt 8 readings/second (PSV mode)
U.S. PATENT D612277
ALL DIMENSIONS IN MM (IN)
PANEL MOUNT
FLANGE
MAX PANEL
THICKNESS
11.7 (0.46)
Includes all effects of linearity, hysteresis, repeatability,
temperature, and stability for one year.
All models indicate vacuum, but vacuum specification applies
to 15 psi, 30 psi, 100 psi, and 300 psi models only.
PORT POSITION
FOR -F4 AND
-RP OPTIONS
Not recommended for continuous use below -14.5 psi.
Refer to XP2i-DP data sheet for gauges that are intended for
continuous high vacuum use.
300 PSI
Non-condensing. No change in accuracy over operating
temperature range. Gauge must be zeroed to achieve rated
specification.
LCD readable in sunlight with bright backlight.
PSV mode is intended for Pressure Safety Valve testing, and
requires software to enable. (ConfigXP is a free download
from XP2i.com.)
SENSOR
DIAPHRAGM
SURFACE 23.6
(0.93)
83.8
(3.30)
(Ø 114.3
(Ø 4.50)
19 (3/4")
WRENCH HEX
7/16-20 MP CPF PORT
43.0
(1.70)
INTERNAL PRESSURIZED VOLUME:
295 mm3 (0.02 in3)
36.1
(1.42)
1/4 INCH MNPT
ADAPTER
29.1
(1.15)
1/4 INCH MBSP
ADAPTER
26.1
(1.03)
18.0
(0.71)
M20X1.5
ADAPTER
32.6
(1.30)
ADDITIONAL SENSOR LENGTH WITH OPTIONAL FITTING ADAPTERS
4533 . L XP2i psi Data Sheet
Page 1 of 5