LR-Cal LCC 100 Pressure Controller (pneumatic)0 pages
Pressure Controller - Calibrator LCC 100 LCC 100
for low pressures / vacuum / diff. pressure
DRUCK & TEMPERATUR Leitenberger GmbH
Bahnhofstr. 33 C D-72138 Kirchentellinsfurt C Germany
Tel.: +49 - 71 21 - 9 09 20 - 0 C Fax: +49 - 71 21 - 9 09 20 - 99
E-Mail: DT-Export@Leitenberger.de C http://www.druck-temperatur.de
Technische Änderungen vorbehalten. Freibleibend C (Rel. 101214) • All technical modifications reserved. Without engagement
Electronic Pressure Controller and Calibrator
Type LCC 100
• generates and measures pressure and vacuum
• pressure up to 1 mbar, 10 mbar, 100 mbar
and 1000 mbar (vacuum max. -600 mbar)
• Uncertainty up to ±0.1% FS (±1 digit)
• Great adjustment accuracy (0.01% FS)
• Chargeable Li-Ion battery
• Generates pressure by pressing a key
The LCC 100 measures and generates pressure and vacuum and
consequently it is more than a mere measuring instrument - it can also be applied as test and calibration instrument
for pressure sensors, pressure switches and pressure gauges. Thanks to internal pressure/vacuum generation no
auxiliary tools are required for operation. An internal Li-Ion battery (chargeable) makes mobile operation very easy.
Parallel operation of mains and battery supply allows for maximum flexibility.
Typical applications:
• Mobile and stationary pressure standard for low pressure and vacuum
• Mobile and stationary of all kind of pressure reading instruments
• Mobile and stationary pressure and vacuum source for calibration purposes
• Leak test
Technical Data:
Type LCC 100-1 LCC 100-10 LCC 100-100 LCC 100-1000 LCC 100-2000
Order-Code: LCC-100-1 LCC-100-10 LCC-100-100 LCC-100-1000 LCC-100-2000
Pressure range: -1...+1 mbar -10...+10 mbar -100...+100 mbar -1000...+1000 mbar -1000...+1000 mbar
Overpressure: 5-fold 5-fold 5-fold 2-fold 2-fold
Uncertainty: ±1 digit ±0.3% FS ±0.1% FS
Linearity: ±1 digit ±0.2% FS ±0.1% FS
Hysteresis: max. 0,1% v.E.
Measurement principle: inductive
Temperature drift of the
internal reference sensor:
Zero point: 0.003% FS / K (0% via zero balance)
Span: 0.03% FS / K
ZERO balance: • automatic (in settable time spans, possible to switch off), or
• manual (by pressing the ZERO button)
Long-term stability of the
internal reference sensor:
0.1% FS per year (typical)
Working temperature range: Working: +10°C...+40°C; Storage: -10°C...+70°C
Usable pressure measuring range: -10...+110%
Adjustment accuracy: 0.01% FS
Adjustment time: depending on volume <5 sec.
Measurement media: Air, non-aggressive and corrosion-free dry gases
Operation modes: • CTRL controlling pressure
• MESS measuring pressure
• AUTO individually definable pressure profile
• Remote-controlled operation (via USB or RS232 interface)
Display: Graphic display (white on blue background)
Interfaces: USB and RS232
Supply: 24 VDC / 1 A, build-in Li-Ion battery (8h typical)
Pressure ports: 6,6 x 11 mm (for flexible hose with 6 mm diameter)
Dimension: Height 102.6 x Width 257 x Dept 271 mm (without handle)
The integrated electr. pump can make max. -600 mbar vacuum. If a
larger vacuum is required, an external vacuum source is needed,
e.g. calibration handpump model 2941.