CLX-Ex0 pages
ft
nnnnWater Quality Measurement Solutions
nnnnA Watts Water Technologies Company
nnnnCC^C-Ex OnLine Residual Chlorine Monitor for hazardous environments II
nnnnThe HF scientific cizx:-Ex OnLine Residual Chlorine Monitor is a rugged, safe, and low
nnnnmaintenance instrument for shipboard ballast water treatments systems that require continuous
nnnnmonitoring and control of residual Total Chlorine levels in hazardous environments. User selectable
nnnncycle times, 4-20mAand RS-485 with Modbus output that can be used to control feed pumps and a
nnnnuser selectable alarm allows complete manual or automatic control of chlorine dosing.
nnnnStandard Features
nnnnOptimized for marine applications
nnnnUse in atmospheres where combustible fuel vapors may be
nnnnpresent
nnnnHousing certified for ATEX, and Zone 1 and Zone 2
nnnnTotal chlorine
nnnnCan be used for Total Chlorine measurements
nnnnAble to operate in ambient temperatures up to 55°C
nnnnLow operating costs
nnnnLow reagent use translates into low operating cost
nnnnSimple operation
nnnnDesigned for unattended operation up to 90 days
nnnnColorimetric DPD chemistry
nnnnCycle times
nnnnCycle times can vary between 110 seconds to 10 minutes (optional 60
nnnnseconds to 10 minutes)
nnnnAlarms
nnnnUser selectable alarm for sample concentration
nnnnBallast water treatment, Drinking water, Wastewater and more
nnnn4-20mA and RS-485 with modbus
nnnnATEX certified
nnnnSuitable for Zone 1 and Zone 2 environments
nnnnUSEPA accepted method
nnnnOrdering Information
nnnnCat. No. Description
nnnn28031
nnnnCLX-Ex OnLine Chlorine Residual Monitor 240VAC for Total Chlorine
nnnncomplete with operating manual.
nnnnReagents must be purchased separately.
nnnnAccessories
nnnn09950 Replacement tubing/cuvette kit
nnnnSpecifications
nnnnRs ncjs. | 0-10 mn/l \j i \j 11 ly/1 |
Cycle Time: | User Selectable: 110 seconds -10 minutes WfJUUIICtl. UU oCUUI IUo - I U IIIIIIUICo |
Accuracy: | 0 - 6 mg/l - ±5% or 0.03 mg/l of CI2 whichever 10 y 1 caici 6-10 mg/l -±10% |
Resolution: | 0.01 mg/l |
ivieinou. | USEPA accepted DPD method of analysis for measuring Total Residual Chlorine |
Standard Outputs: | 4-20 mA and RS-485 with Modbus |
User Alarm: | User selectable alarm for sample concentration |
Operating Temperature: | 0° - 55°C (32° to 131°F) |
Sample Pressure: | 0.3 bar -10.3 bar (5 -150psi) |
Compressed Air Pressure: | 5.5 - 7 bar (80 -101.5 psi) |
Enclosure: | IP66 |
Power: | 240 VAC (130 to 240VAC), 47-63 Hz, 250VA |
Certifications: | CE, ATEX |
Flow Rate to Waste: | 200-400ml / min |
Shipping Dimensions | 558mm X 662 mm X 400 mm (22" x 24 112" x 15 3/4") |
Shipping Weight | 38 kg (83 lbs.) Reagents are Shipped Separately |
Actual Weight | 26.3 kg (58 lbs.) |
Installed Dimensions | 19.7" x 18.7" x 9.4" (50cm x 48cm x 23.8cm) |
Specifications subject to change without notice.
nnnnReagent Kits
nnnnCatalog number
nnnnTotal Chlorine reagent for CLX-EX 12 month supply
nnnn09991
nnnn(based on estimated BWTS operational cycles)
nnnnHF scientific 3170 Metro Parkway, Ft. Myers FL 33916 e-mail: hfinfo@watts.com
nnnnph: (239)-337-2116 • fax: (239) 332-7643 • toll free: 888-203-7248 www.hfscientific.com
nnnnRev. 11-17-2011
"