MA600MHI0 pages
MA600MHI Series
High Isolation, 6W
Medical Approved
DC/DC Converters
Electrical Specications
Specications typical @ +25°C, nominal input voltage & rated output current, unless otherwise noted. Specications subject to change without notice.
Key Features:
Input
Parameter
• 4 kVAC rms Isolation
• Reinforced Insulation
• 2 µA Leakage Current Max
• Wide 2:1 Input Range
• Compact DIP Case
• Single & Dual Outputs
• Meets EN55022
• 700 kH MTBF
• EN 60950 Approved
• EN 60601 Approved
Input Start Voltage
Under Voltage Shutdown
Input Filter
Leakage Current
Short Circuit Input Power
Conditions
Min.
Typ.
Max.
12 VDC Input
7.0
8.0
9.0
24 VDC Input 13.0
15.0
18.0
48 VDC Input 30.0
33.0
36.0
12 VDC Input
8.5
24 VDC Input
16.0
48 VDC Input
34.0
(Pi) Filter (Complies with EN55022 Class A)
240 VAC, 60 Hz
2
3,000
Units
Conditions
• 6W Output Power
Units
%
%
%
%
VDC
VDC
µA
mW
Output
Parameter
Output Voltage Accuracy
Output Voltage Balance
Line Regulation
Load Regulation
Ripple & Noise (20 MHz), See Note 1
Ripple & Noise (20 MHz)
Ripple & Noise (20 MHz)
Output Power Protection
Transient Recovery Time, See Note 2
Transient Response Deviation
Temperature Coefcient
Output Short Circuit
Min.
Dual Output , Balanced Loads
Vin = Min to Max
Iout = 25% to 100%
5V Output
All Other Outputs
Over Line, Load & Temp.
Typ.
±0.5
±0.5
±0.3
±0.5
75
100
Max.
±1.0
±2.0
±0.5
±1.0
100
150
180
25
120
300
±3.0
±0.02
Continuous (Autorecovery)
25% Load Step Change
500
±6.0
±0.05
mV P - P
mV P - P
mV rms
%
µSec
%
%/ºC
General
Parameter
Isolation Voltage, Rated
Isolation Test Voltage
Reinforced Insulation Working Voltage
Isolation Resistance
Isolation Capacitance
Switching Frequency
Conditions
60 Seconds
Flash Tested For 1 Sec
500 VDC
100 kHz, 1V
Min.
4,000
6,000
300
1,000
Typ.
Max.
7
150
13
Units
VAC rms
VPk
VAC
M⍀
pF
kHz
Environmental
Parameter
Operating Temperature Range
Operating Temperature Range
Storage Temperature Range
Cooling
Humidity
Conditions
Min.
Typ.
-40
-40
-50
Free Air Convection
RH, Non-condensing
Ambient
Case
Max.
+75
+95
+125
Units
ºC
ºC
ºC
95
%
Physical
MicroPower Direct
292 Page Street
Suite D
Stoughton, MA 02072
USA
T: (781) 344-8226
F: (781) 344-8481
E: sales@micropowerdirect.com
W: www.micropowerdirect.com
Case Size
Case Material
Weight
1.25 x 0.80 x 0.47 Inches (31.8 x 20.3 x 12.0 mm)
Non-Conductive Black Plastic (UL94-V0)
0.63 Oz (18.0g)
Reliability Specications
Parameter
MTBF
Safety Approvals
Conditions
Min.
Typ.
Max.
MIL HDBK 217F, 25ºC, Gnd Benign
700
UL 60601, UL 60950, EN 60601, EN 60950
Units
kHours
Absolute Maximum Ratings
Parameter
Input Voltage Surge (1 Sec)
Lead Temperature
Internal Power Dissipation
Conditions
12 VDC Input
24 VDC Input
48 VDC Input
1.5 mm From Case For 10 Sec
All Models
Min.
-0.7
-0.7
-0.7
Typ.
Max.
25.0
50.0
100.0
260
2,500
Caution: Exceeding Absolute Maximum Ratings may damage the module. These are not continuous operating ratings.
www.micropowerdirect.com
Units
VDC
ºC
mW