DRT-15CE Portable Turbidimeter0 pages
ft
nnnnWater Quality Measurement Solutions
nnnnA Watts Water Technologies Company
nnnnDRT-15CE PORTABLE TURBIDIMETER
nnnnDesigned to provide the rugged portability needed in the field with the accuracy
nnnndemanded in the laboratory, the EPA accepted DRT-15CE has become a "must have"
nnnnfor anyone monitoring turbidity on the go. The integral carrying case holds everything
nnnnneeded for field operation. With resolution of 0.01 NTU and an extended range to
nnnn1,000 NTU, the DRT-15CE is a perfect fit for both field and laboratory use.
nnnnStandard Features
nnnn• Rugged Portable Carrying Case
nnnnEnsures durability and convenience no matter where your laboratory
nnnnhappens to be
nnnn• Completely Self-Contained
nnnnProvides easy access to everything you need to take a measurement.
nnnn• Three User Selectable Ranges
nnnnProvides accuracy in a wide variety of applications from 0 to 1000 NTU.
nnnn• 20 Hour Rechargeable Battery
nnnnAlways be ready to go with up to 20 hours of continuous operation when
nnnnfully charged.
nnnnOptional Features
nnnn• Automobile Adapter
nnnnCharge and operate the DRT-15CE from the cigarette lighter in your
nnnnautomobile.
nnnn• fmCd Calibration Standards
nnnnQuick and easy calibration across all ranges without the need to mix
nnnnFormazin.
nnnn• Formazin Stock Solution Kit
nnnnContains everything you need to calibrate the DRT-15CE with Formazin, the
nnnnprimary calibration standard.
nnnnOrdering Information
nnnnCat. No. Description
nnnn19057
nnnn19058
nnnnDRT-15CE Portable Turbidimeter, 120 VAC
nnnnDRT-15CE Portable Turbidimeter, 240 VAC
nnnn* The DRT-15CE exceeds performance criteria specified by US EPA
nnnnmethod 180.1 for NTU measurement.
nnnnAccessories
nnnn19045
nnnn39071
nnnn50051
nnnn50052
nnnn12V Automobile Adapter to operate & charge the DRT-15CE from a
nnnncigarette lighter plug.
nnnnfroCdCalibration Kit, (0.02,10.00, & 1000 NTU)
nnnnCuvettes, 28 x 70mm, 3 pack
nnnnCuvettes, 28 x 70mm, 10 pack
nnnnSpecifications
nnnnRange: | 0-10 NTU, 0 -100 NTU, 0 - 1000 NTU |
Accuracy: | 0-10 NTU, ± 1% of full scale 0-100 NTU, ± 5% of full scale 0-1000 NTU, ± 10% of full scale |
Resolution: | 0.1 % of selected range |
Response Time: | Immediate |
Sample Size: | 25 ml (16ml minimum) |
Battery: | 6V gel-cell rechargeable with 20 hours of continuous use when fully charged. |
Recharge Time: | 8 hours to full charge |
Operating Temperature: | 32°-122°F(0°-50°C) |
Dimensions: | 11 x 8 x 5 in. (28 x 20 x 14 cm) |
Weight: | 5 lbs. (2.27 kg) |
Standards Compliance: | USEPA |
* The DRT-15CE exceeds performance criteria specified by US EPA
nnnnmethod 180.1 for NTU measurement.
nnnnSpecifications subject to change without notice.
nnnn3170 Metro Parkway, Ft. Myers FL 33916 e-mail: hfinfo@watts.com
nnnnph: (239)-337-2116 • fax: (239) 332-7643 • toll free: 888-203-7248 www.hfscientific.com
nnnn8-04-2010 Rev 1
"