GS01E06K02-00E0 pages
<> <>
General
Specifications
Model AM012
Magnetic Flow Converter
GS 1E6K2-E
OUTLINE
AMO12 is a signal generator used to check the calibration
and operation of AM11 MAGNETIC FLOW CONVERTERS,
AM 12 MAGNETIC FLOW CONVERTERS,
AM100A/200A-INTEGRAL TYPE MAGNETIC FLOWMETER,
YMA11 MAGNETIC FLOW CONVERTER, COMPACT
YEWMAG MAGNETIC FLOWMETER (style A and style B)
and ADMAG AE MAGNETIC FLOWMETER.
As the excitation load is build-in, the calibration and
operation of a converter can be checked without a detector.
Span can be set within the range of flow rates 0 to 10 m/s.
Output signal correponding to setting span can be set to 0,
25, 50, 75, 100% with the setting dial.
Power Code
Power plug is attached as described below.
STANDARD SPECIFICATIONS
Case:
Material:Polyester resin
Size:110 (height) x235(width) x364(rear)mm.
Color: Light grayish green.
Connecting Cable: Power cord for AM012.
Signal cable.
Excitation cable.
Setting Ranges:
Span Setting Dial: 0 to 10 m/s (minimum span 0.01 m/s).
Output Setting Dial: CHECK, 0, 25, 50, 75, 100%.
Accuracy:
0.1 m/s or more: +0.04% of set value.
0.1 m/s or less: Flow rate of +0.04 mm/s.
Power: 90 to 132 V AC or 180 to 264 V AC, 47 to 63 Hz
Ambient Temperature: 0 to 40
Weight: 4.2kg
Model
Suffix code
Description
AM012
..........................
Power
code
-1....................... 100 V version
-2....................... 100 V version
-3....................... 200 V version
Style code
Optional
specification
Calibrator for Magneric Flowmeter
* A.............. Style A
/CYM With Cable for compact YEWMAG
With Junctionbox for ADMAG AE
/AE
and ADMAG CA
With Junctionbox for ADMAG SE,
/SE
and special cable
OPTIONAL SPECIFICATION
Cable for Compact YEWMAG
For the calibration of Compact YEWMAG.
With the connector for signal and excitation cable.
Option code /CYM
Junctionbox for ADMAG AE and ADMAG CA
When using the AM012 calibrator with the ADMAG AE
and ADMAG CA, remove only the amplifier from the
converter and put it in the dedicated junctionbox for
calibration. Use this item when only the junctionbox for
AM012 is to be procured.
Option code /AE
Junctionbox for ADMAG SE
When using the AM012 calibrator with the ADMAG SE,
remove only the amplifier from the converter and put it in
the dedicated junctionbox for calibration. Use this item
when only the junctionbox for AM012 is to be procured.
Option code /SE
ACCESSORIES
Power code ................................................................... 1set
Fuse ( 0.2A) ................................................................ 1set
Signal cable .............................................................. 1set
Excitation cable ....................................................... 1set
Junction box (/AE, /SE) ................................................ 1set
Cable for ADMAG SE(/SE) .............................................. 1set
Cable for Compact YEWMAG(/CYM) ........................... 1set
Yokogawa Electric Corporation
2-9-32, Nakacho, Musashino-shi, Tokyo, 180 Japan
Tel.: 81-422-52-4443 Fax.: 81-422-52-2018
GS 1E6K2-E
© Copyright June 1989(YK)
4th Edition: Dec. 1997(YK)