SG750 Stack Gas Analyzers0 pages
EXA IR
General
Specications
Model SG750 Stack Gas Analyzer
GS 11G04G01-01E
GENERAL
The SG750 Stack Gas Analyzer consists of an infrared gas analyzer, a zirconia oxygen analyzer, and a
sampling unit.
The SG750 can simultaneously measure up to ve
components: sulfur dioxide (SO2), nitrogen oxides
(NOX), carbon monoxide (CO), carbon dioxide (CO2)
and oxygen (O2).
FEATURES
1. Simultaneously measures up to ve gas components
The SG750 uses a combination of an infrared gas
analyzer(s) and a dedicated zirconia or paramagnetic oxygen analyzer to simultaneously determine
concentrations of up to ve gas components - NOX,
SO2, CO, CO2 and O2.
2. Minimizes the effects of interference from other gas
components
The use of interference compensation in the analyzer virtually eliminates the effects of interference
from other gas components.
3.Maintenance can be performed from front side, thus
saving the installation space. Unitized structure
of the analyzing block and gas sampling module
enables better maintenance.
4. Provides a wealth of functions
The SG750 provides a variety of standard functions,
including reliability-enhancing self-diagnosis, auto
matic calibration, correction to O2 values, averaging,
high and low limit alarms, and more.
5. Wide dynamic range
The SG750 is highly sensitive with a wide dynamic
range which can be switched up to 1:25 and which
can be changed arbitrarily.
SAMPLING SYSTEM CONFIGURATION
Example 1: Five-Component Gas Sampling System Configuration
SO2 1st range: less than 500 ppm and sample gas pressure range is -1 to 5 kPa or -3 to 3 kPa, Standard type.
<1> Probe
<8> Switch
Box
Air Inlet
Teflon
Tube,
Φ10/Φ8
re
mo
or
˚
15
*1
<4> Sampling Module
<7> 3-Way
Solenoid Valve
<3> Pump
Sample
Gas Inlet,
Rc 3/8
*1
Needle
Valve
<2>
Gas
Conditioner
Membrane Filter
2-Way
Solenoid Valve
Flow Checker
NO2/NO
Converter
30 kPa
0.5L/min
Atmospheric
Air
Membrane
Filter
Electric
gas cooler
Solenoid
Valve
Infrared
Analyzer
CO, CO2
Infrared
Analyzer
NO, SO2
ZERO
NO
SO2
CO2
CO
O2
Pressure Regulator
Air
NO
/N2
SO2
/N2
CO2
/N2
CO
/N2
O2
/N2
<6>
Zirconia
O2
Analyzer
Drain
Filter
Regulator
30 kPa
<5> Standard Gas
Instrument Air Inlet
Yokogawa Electric Corporation
2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 Japan
Tel.: 81-422-52-5617 Fax.: 81-422-52-6792
Outlet
GS 11G04G01-01E